1055018 | 透析袋md18(5000) | 压平宽度18mm,直径11.5mm,单位长度体积1.1 ml/cm | 现货 |
1055030 | 透析袋md30(5000) | 压平宽度30mm,直径19.1mm,单位长度体积2.86 ml/cm | 现货 |
1055040 | 透析袋md40(5000) | 压平宽度40mm,直径25.5mm,单位长度体积5.1 ml/cm | 现货 |
Cellu·Sep H1的透析袋纯度比T-Series 系列产品高,采用EDTA处理技术,不含重金属离子,湿型,储存于特制的塑料管中;大部分是0.5mx10的包装,也有5m和10m的包装。
The regenerated cellulose in Cellu·Sep® is derived from cotton: Cotton linters are dissolved in a solution and spread into flat sheets or extruded into tubes. The material is treated with glycerin to prevent the pores from collapsing and air dried at a certain temperature and pressure to form a rigid membrane. Cellu·Sep regenerated cellulose membrane has a symmetric pore structure which allows small molecules to migrate in either direction, making it ideal for experimental purposes.
即用型透析袋包括美国光谱医学(spectrumlabs)和美国迈伯伦(MFPI cellu.sep系列)公司生产的透析袋,处理方法简单,但是客户往往因不清楚而采取了繁琐且不正确的处理方法,本文重点介绍即用型透析袋处理方法和使用方法。
一、即用湿型透析袋的处理和使用方法:
1、储存条件:
透析膜浸泡在储存液中,封放置于4-37 °C之间环境,有效期3年。2、透析袋使用方法:
①把透析袋剪成适当长度(10cm左右)的小段。
②用去离子水浸泡10分钟左右,再用去离子水膜内冲洗3次。
③不使用的透析膜放回储存液中,密封保存,接触透析膜过程中必须戴手套。
④使用时,一端用下沉透析袋夹子夹紧,灌满水后,用手指适当加压,检查不漏,方可装入样品。通常要留三分之一至一半的空间,以防透析过程中,袋外的水和缓冲液过量进入袋内将袋涨破。装完样品后,用上浮夹子夹紧袋口,可在袋内放一玻璃珠,以使透析袋处于悬浮状态,即可进行透析(详见透析袋装置简图)。为了加快透析速度,除多次更换透析液外,还可使用磁力搅拌器。透析的容器要大一些,可以使用大烧杯、大量筒和塑料桶。
3、透析后去除透析液方法:
透析后,透析袋中除样品外还有一定量透析液,我们推荐使用美国光谱医学生产的透析脱水剂(Spectra/Gel Absorbent)。
二、即用干型透析袋的处理和使用方法:
1、储存条件:
干型透析膜表层涂有甘油保护层,封放置于4-37 °C之间环境,有效期5年。
2、透析袋使用方法:
①把透析袋剪成适当长度(10cm左右)的小段。
②用60℃去离子水中水浴30分钟-1小时,再用去离子水膜内冲洗3次。
③使用时,一端用下沉透析袋夹子夹紧,灌满水后,用手指适当加压,检查不漏,方可装入样品。通常要留三分之一至一半的空间,以防透析过程中,袋外的水和缓冲液过量进入袋内将袋涨破。装完样品后,用上浮夹子夹紧袋口,可在袋内放一玻璃珠,以使透析袋处于悬浮状态,即可进行透析(详见透析袋装置简图)。为了加快透析速度,除多次更换透析液外,还可使用磁力搅拌器。透析的容器要大一些,可以使用大烧杯、大量筒和塑料桶。
3、透析后去除透析液方法:
透析后,透析袋中除样品外还有一定量透析液,我们推荐使用美国光谱医学生产的透析脱水剂(Spectra/Gel Absorbent)